POSITIVE: One red line appears in the control region (C). No line appears in the test region (T). This positive result indicates the drug concentration is above the detectable level. Please refer to the Illustration above.
NEGATIVE: Two lines appear. One red line should be in the control region (C), and another red or weak pink line adjacent appears in the test region (T). This negative result indicates the drug concentration is below the detectable level.
*NOTE: The shade of red in the test line region (T) may vary, but it should be considered negative whenever there is even a faint pink line. Please refer to the Illustration above.
INVALID: Control line fails to appear. Incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test using a new test device. If the problem persists, discontinue using the lot immediately and contact your local distributor. Please refer to the Illustration above. The results can be read for a period of up to 10 minutes. DO NOT INTERPRET RESULTS AFTER 10 MINUTES as they will no longer be valid.
Unscrew the Interlocking Lid. This process will break the Interlocking Seal and open the Interlocking Lid, exposing the Upper Collection Chamber.
Collect the urine specimen from the Test Chamber with a syringe and needle to perform the confirmatory test.
Insert the pointed Puncture Tool provided into the Puncturing Slot (follow indication arrow) located in the Upper Collection Chamber and the Intermediate Chamber. This will give you access to the Test Chamber.
After the Interlocking Lid is screwed back on, the confirmed positive specimen can be stored/frozen for evidentiary use.
Manually record donorʼs biographical information, as well as the test device serial number.
When ready to begin testing, remove the test device from the sealed pouch. Do not tear open the pouch until ready to begin testing.
Fill in the requested information on the device (top and side) using a non-erasable ballpoint pen or permanent marker.
Issue the device to the donor and have them void directly into the Upper Collection Chamber. Ensure the urine specimen is below the maximum fill line.
The device must be returned to you. Place the Interlocking Lid loosely on top, ensuring the Anti-Locking Protective Ring is not removed from the device at this point.
Check the device to reconfirm the serial number matches the one issued earlier to the donor as recorded in Step 1.
Open the Interlocking Lid and pour off any excess urine above the maximum level indicator. Remove the Anti-Locking Protective Ring and level the Interlocking Lid to the device evenly.
Twist the Lid until tightly secured. Be sure the Interlocking Lid is locked into the Interlocking Seal. DO NOT ATTEMPT TO REOPEN THE LID, AS THE INTERLOCKING SEAL WILL BREAK.
Allow the urine specimen to flow down to the Test Chamber. Wait FIVE MINUTES to read/record results from the window showing the test results. DO NOT INTERPRET RESULTS AFTER 10 MINUTES AS THEY WILL NO LONGER BE VALID.
Also check the Temperature Strip on the deviceʼs outside wall to ensure the sample is fresh/sufficiently thawed. The normal temp range should be 37°C/98.6°F+/-0.61°C/1°F.
If there are any positive results, generate a laboratory confirmation test request. If the positive field test samples are used for confirmatory tests, they must be conducted within the specific storage time and conditions prescribed for the drug being tested.
Copyright © 2024 Global Risk Group. All rights reserved. Designed by UNIweb